Improved segmented modified Look-Locker inversion recovery T1 mapping sequence in mice
Revista : PLOS OneVolumen : 12
Número : 11
Páginas : 13pp
Tipo de publicación : ISI Ir a publicación
Abstract
Object
To develop and evaluate a 2D modified Look-Locker (MOLLI) for high-resolution T1 mapping in mice using a 3T MRI scanner.
Materials and methods
To allow high-resolution T1 mapping in mice at high heart rates a multi-shot ECG-triggered 2D MOLLI sequence was developed. In the proposed T1 mapping sequence the optimal number of sampling points and pause cardiac cycles following an initial adiabatic inversion pulse was investigated in a phantom. Seven native control and eight mice, 3 days post myocardial infarction (MI) after administration of gadolinium were scanned. Two experienced readers graded the visual T1 map quality.
Results
In T1 phantoms, there were no significant differences (<0.4% error) between 12, 15 and 20 pause cardiac cycles (p = 0.1, 0.2 and 0.6 respectively) for 8 acquisition cardiac cycles for 600bpm in comparison to the conventional inversion recovery spin echo T1 mapping sequence for short T1s (<600 ms). Subsequently, all in-vivo scans were performed with 8 data acquisitions and 12 pause cardiac cycles to minimize scan time. The mean native T1 value of myocardium in control animal was 820.5±52 ms. The post-contrast T1 measured 3 days after MI in scar was 264±59 ms and in healthy myocardium was 512±62 ms. The Bland-Altman analysis revealed mean difference of only -1.06% of infarct size percentage between T1 maps and LGE.
Conclusions
A multi-shot 2D MOLLI sequence has been presented that allows reliable measurement of high spatial resolution T1 maps in mice for heart rates up to 600bpm.